Roxar Analyser Transmitter 6 Monthly Instrument Preventive Maintenance AT
Note: Prior to
Commencing this task obtain the information required is as follows:
A) Span is 4 to
20 milliamps.
B) Relay actuation
Set point.
C) Cut
calibration (grind out value).
D) Temperature
coefficient.
All these
values may be entered with the three buttons located on the front of the water
cut monitor.
TASKS TO BE
PERFORMED:
1. Annual
Function Check and Calibration of Invalco Analyser Transmitter.
2. Inspect
Analyser Transmitters for any physical damage.
3. Ensure
Transmitter is securely mounted.
4. Inspect
electrical cable and fitting for any physical damage.
5. Check the
span setting of the unit is 4 to 20 Milli Amps.
6. Depress all
three buttons on the front of the WCM to enter the programming mode.
7. Press enter
to accept the delay and to proceed to the next display.
8. The display
will now read Undo changes Y/N with the N flashing for NO.
9. The left
arrow will toggle the cursor between YES and NO.
10. DO NOT
SELECT YES UNLESS YOU INTEND TO RECONFIGURE THE UNIT COMPLETELY.
11. Selecting
YES would undo all information entered during the programming session and
restore the unit to all conditions prior to entering the programming mode.
12. Leave the
selection at N.
13. Press ENTER
to accept all of the settings and proceed to the next display.
14. The display
will now read programme done.
15. Press ENTER
TO LEAVE THE PROGRAMME MODE.
16. The unit
should now be reading the same value as the grind out unless the water cut has
changed.
17. Calibration
is complete for a single cut.
TAKE SAMPLE FOR
LABORATORY ANALYSIS
Test Equipment:
Laboratory Sample bottle minimum 1 litre to achieve a reasonable representative
sample. Karl Fischer Coulomb-meter or Equivalent.
Procedure:
a) Ensure that
Flow has stopped through the desired stream.
b) Once flow
has ceased through the stream, it must be left for 4 - 6 hours to stabilise and
any water content to separate.
c) A sample of
the Oil must be taken from the Highest point available and as near to the
monitor as feasibly possible and taken to the Chemists Laboratory for
analysis.(Preferably in the morning).
d) Obtain the
BS&W results from this sample from the Chemist's Laboratory.
e) Enter the
results of the sample into the unit following the Manufacturer data on Setup
and calibration.
f) Allow the
temperature to rise or fall by at least 3 Deg C and perform steps c) though to
e) again, the unit will auto compensate for the temperature change to establish
a constant which will distinguish between temperature effect on the probe and
actual contamination by water and sediment.
g) Return the
unit to normal operation and resume flow through the stream.
18. Sign-off
Work Permit and return to operations.
19. Report all
findings to supervision and document in CMMS.